A high-sensitivity single-frequency method for low-loss thin dielectrics. Using the Keysight PNA-X Vector Network Analyzer with a QWED split post dielectric resonator, iEMSL measures dielectric constant and loss tangent at 10 GHz, including measurements up to 100 °C. In collaboration with QWED.
10 GHz
Single frequency
±2×10⁻⁵
tanδ resolution
To 100 °C
Temperature option
5 business days
Standard turnaround
What Is Split Post Dielectric Resonator (SPDR) Testing?
Split post dielectric resonator (SPDR) testing measures the dielectric constant (Dk) and loss tangent (Df) of a flat sheet or thin film at a single frequency of 10 GHz. The Intelligent Electromagnetic Sensor Laboratories (iEMSL) at Texas A&M University offers this split post dielectric resonator testing service at room temperature and up to 100 °C.
Samples are flat sheets at least 45 mm on each side, no wider than 90 mm and no thicker than 0.95 mm. Films on a substrate are accepted, with a bare substrate sample as well. The method supports IPC 2.5.5.15 and IEC 61189-2-721. External pricing starts at $300 per sample at room temperature; measurements up to 100 °C and film-on-substrate samples are $500 per sample. SPDR is not for broadband data, because it measures at one frequency only, so a material needed across a band belongs on the FPOR from 10.2 to 52.2 GHz. It is also not for thick or machined solids, liquids or powders.
Why the SPDR
It is the right method for thin sheets, films and film-on-substrate coatings that need a loss figure you can trust, and it is the only iEMSL method that runs to 100 °C. It is the wrong method if you need more than one frequency: the SPDR measures at 10 GHz and nowhere else, so a material whose Dk you need across a band belongs on the FPOR from 10.2 to 52.2 GHz or on WR-90 across the X-band. It is also the wrong method for anything you cannot present as a flat sheet 45 to 90 mm across and under 0.95 mm thick: machined or thick solids go to WR-90, liquids, powders and lossy materials to the DAK-TL2, and anything below the microwave bands to the parallel plate.
Ultra-high sensitivity
Loss tangent resolution to ±2×10⁻⁵, ideal for extremely low-loss ceramics and specialized polymers used in high-Q structures.
Film-on-substrate
Extracts a coating’s properties from the shift in resonant frequency and Q-factor between the bare and coated substrate.
Thin sheets and films
Characterizes free-standing dielectric sheets and thin films at a precise single frequency.
High temperature
The only iEMSL standard method offered up to 100 °C, supporting IPC 2.5.5.15 and IEC 61189-2-721.
Film-on-substrate measurement
We extract a coating’s properties by measuring the bare substrate first to set a baseline, then the coated substrate, and tracking the shift in resonant frequency and Q-factor. Send a bare reference piece of the same substrate, from the same lot, alongside your coated sample. Without the reference, the film and substrate cannot be separated.
Technical specifications
| Frequency | 10 GHz (single frequency) |
|---|---|
| Dielectric constant | Δε/ε = ±(0.0015 + Δh/h) |
| Loss tangent | ±2×10⁻⁵ or ±0.03 × tanδ |
| Temperature | Room temperature to 100 °C |
| Standards | IPC 2.5.5.15, IEC 61189-2-721 |
Sample requirements
Lateral size: 45-90 mm (circular or square)
Thickness: 15 μm – 0.95 mm
Form: planar, with clean edges
For film-on-substrate work, include a bare reference of the same substrate from the same lot.
What to send us
- One planar sample per measurement, 45-90 mm (circular or square), 15 μm – 0.95 mm thick
- For film-on-substrate: a bare substrate reference of the same lot
- Whether high-temperature data (to 100 °C) is required
Pricing
$300 per sample
High-temperature measurement up to 100 °C is $500 per sample. Standard turnaround 5 business days, with rush service available.
Texas A&M researchers receive 50% off listed prices.
What you receive
- Extracted Dk and Df
- Tables versus temperature, where applicable
- Raw data (CSV) including resonance and Q data on request
- Technical report on request (additional charge)
View our example measurements. See also: WR-90, FPOR, parallel plate, all material characterization services.
Frequently Asked Questions
What do I send for a film-on-substrate measurement?
Send the coated sample plus a bare reference piece of the same substrate from the same lot. We measure the bare substrate first to set a baseline, then the coated sample, and extract the film from the shift in resonant frequency and Q-factor. Without the reference, the film and substrate cannot be separated.
Can you measure at high temperature?
Yes, the SPDR is the only iEMSL standard method offered up to 100 °C, priced at $500 per sample.
Is the SPDR single-frequency?
Yes, it measures at 10 GHz. If you need data across a frequency band, use the FPOR (10.2-52.2 GHz).
What sample size and thickness do you need?
A planar sample 45-90 mm across (circular or square), 15 μm to 0.95 mm thick, with clean edges.
How low a loss tangent can you resolve?
Loss tangent resolution is ±2×10⁻⁵ (or ±0.03 × tanδ, whichever is greater), among the most sensitive methods available for low-loss thin dielectrics.
Can you handle ITAR-controlled or export-controlled work?
Yes. iEMSL accepts ITAR-controlled projects under a Technology Control Plan, administered with the Texas A&M export control office. Tell us your control requirements when you request a quote, before shipping any samples.
Ready to characterize your thin film?
From low-loss substrates to coated films, get sensitive 10 GHz data.
